text.skipToContent text.skipToNavigation
Carbon Dioxide Safety Monitor AX60

Analox Sensor Technology Ltd.

Product attributes

Select which additional attributes you want to display in the table below
Product Description Packaging Qty Price € Availability Add to Cart
LLG06284264 Carbon Dioxide Safety Monitor AX60, Additional CO2 Alarm (Max 8 per Central Display Unit) Additional CO2 Alarm (Max 8 per Central Display Unit) 1 pz.
Item price: 133,00
LLG06284263 Carbon Dioxide Safety Monitor AX60, Additional CO2 Sensor (Max 4 per Central Display Unit) Additional CO2 Sensor (Max 4 per Central Display Unit) 1 pz.
Item price: 313,20
LLG06284262 Carbon Dioxide Safety Monitor AX60, AX60 CO2 Safety Monitor AX60 CO2 Safety Monitor 1 pz.
Item price: 642,20
LLG06284265 Carbon Dioxide Safety Monitor AX60, Optional Sensor Protector Kit: Protector bar, splashguard and fixing kit Kit opzionale Protezione Sensore: Barra di protezione, para-spruzzi e kit di fissaggio 1 pz.
Item price: 58,60

Legend

Available

Available in 15 days

Available in 30 days

Available in more than 30 days

Carbon Dioxide Safety Monitor AX60
Gas warning device for precise measurement of the amount of CO2 in the air. 3 alarm settings factory set: Time-weighted average (TWA) alarm (5000 ppm; 0.5% CO2), Low alarm (15,000 ppm; 1.5% CO2) and High alarm (30,000 ppm; 3% CO2)
-No calibration required (unless local regulations state otherwise)
-Very long life sensor can last up to 15 years
-Up to 4 x CO2 sensors and 8 x CO2 alarms can be used per Central Display Unit
-Supplied with a relay on each alarm which can be used to trigger ventilation, extraction or a building management system
-Easy installation and maintenance can be conducted by the user
-Additional (quick connect) CO2 sensors and alarms available if more than one entrance to the laboratory or more than one risk area
-Assists in compliance with Exposure Limit regulations such as EH40

Scope of delivery: wall-mounted central display unit, CO2 sensor and CO2 alarm which can provide an early warning of a CO2 leak.
Specifications:
Range :
Accuracy : 0.1 to 5 % CO2
+/- 5 % of alarm setpoint
Alarm intensity : 88 dbA
Monitor Display : 16 digit 2 line LCD in ppm
TWA Alarm : 5,000 ppm (0.5 %) over 8 hours
CO2 Alarms : 15,000 ppm (1.5 %) and 30,000 ppm (3 %)
Operating Temperature : -5 to 50 °C
Central Unit Dimensions : 175 x 110 x 75 mm
Sensor & Alarm Dimensions : 72 x 170 x 45 mm
IP Rating : IP55 (IP54 Central Display Unit)
Warranty : 5 years for electronics and sensor